High Resolution Spectra:

O3 Spectra (2001)

O3 Spectra (2011)

BrO Spectra

NO2 Spectra

OClO Spectra

OBrO Spectra

OIO Spectra

IO Spectra

I2 Spectra

Glyoxal Spectra

CH4 spectra near 1.6 um

CH4 data

Absolutely calibrated ozone absorption cross-sections for 11 temperatures (193K – 293K in steps of 10K) for UV-NIR spectral region

More details:

- Spectroscopy of Europe, vol. 23, 6, 2011, pp.14-17
» http://www.spectroscopyeurope.com/articles/55-articles/3082-new-broadband-high-resolution-ozone-absorption-cross-sections

-AMT

» http://www.atmos-meas-tech.net/7/609/2014/amt-7-609-2014.html

» http://www.atmos-meas-tech.net/7/625/2014/amt-7-625-2014.html

» Download the ASCII data, version 25.07.2012, zip, total size 12Mb [12.242 KB]

» Download temperature coefficients between 300 and 370 nm from a quadratic fit [438 KB]

Version 22.07.2013: Fast Fourier Transform Filter applied to the initial data in the region 213.33 -317 nm

» Download the ASCII data, version 22.07.2013, zip [7.910 KB]

Please, check for updates! Measurements with dual channel FTS are ongoing for the IR and region around 380 nm.

In case of any questions please email Dr. Victor Gorshelev: » gorshelev@iup.physik.uni-bremen.de

Source: IUP, MolSpec Lab, Serdyuchenko A., Gorshelev V, Weber M., Chehade W., Burrows JP.
 
Spectrometer: Echelle Spectrometer ESA 4000 and Bruker HR 120 FTS  
Resolution (FWHM):  0.02 nm below 290 nm
1 cm-1 between 290 nm and 350 nm
0.02 nm between 350 nm and 450 nm
1 cm-1 between 450 nm and 1100 nm
 
Grid:  interpolated on grid 0.01 nm   
Cell: Double jacket quartz cell, thermo-insulated, pre-cooler, cryogenic cooling  
Absolute calibration: using pure ozone pressure  
Uncertainty for absolute measurements at 50 mbar and 193 - 293K and path length 270 cm:    
Systematic uncertainty Ozone impurity:
oxygen purity 99.995%
leaks <0.1%.
Accuracy of pressure sensors: 0.04% (0.02 mb)
Offset of the internal temperature sensors: <0.3-0.5% (1K)
Cell length: <0.07% (<1 mm)
Total: <0.5%, depending on temperature
 
Statistical uncertainty Ozone decay <1%
Fluctuations of pressure: < 0.08% (<0.04mb)
Fluctuations of temperature: <0.1 - 0.16% (<0.3K)
Cooling efficiency: <0.3 - 1.6% (1-3K)
Source intensity drift and white noise 0.1 - 70%, depending on spectral region
 
Concatenated spectra parameters:    
Spectral regions: Lightsource stability: Optical density limits:
213–290 nm 0.5% 0.5-2
290–310 nm 2% 0.1-2
310–340 nm 1% 0.1-2
340–450 nm 1% 0.05-1
450–750 nm 0.2% 0.5-2
750–1100 nm 0.2% 0.001-0.1
     
column 1: vacuum wavelength (nm)  
column 2: cross section (cm2/molecule) 293k
column 3: cross section (cm2/molecule) 283k
column 4: cross section (cm2/molecule) 273k
column 5: cross section (cm2/molecule) 263k
column 6: cross section (cm2/molecule) 253k
column 7: cross section (cm2/molecule) 243k
column 8: cross section (cm2/molecule) 233k
column 9: cross section (cm2/molecule) 223k
column 10: cross section (cm2/molecule) 213k
column 11: cross section (cm2/molecule) 203k
column 12: cross section (cm2/molecule) 193k