Source:
|
IUP, MolSpec Lab, Serdyuchenko A., Gorshelev V, Weber M., Chehade W., Burrows JP; Submitted to the JQSRT on 25.07.2012: High spectral resolution ozone absorption cross-sections: Part I. Measurements and Analysis High spectral resolution ozone absorption cross-sections: Part II.Temperature dependence
|
|
|
Spectrometer:
|
Echelle Spectrometer ESA 4000 and Bruker HR 120 FTS
|
|
|
Resolution (FWHM):
|
0.02 nm below 290 nm 1 cm-1 between 290 nm and 350 nm 0.02 nm between 350 nm and 450 nm 1 cm-1 between 450 nm and 1100 nm
|
|
|
Grid:
|
interpolated on grid 0.01 nm
|
|
|
Cell:
|
Double jacket quartz cell, thermo-insulated, pre-cooler, cryogenic cooling
|
|
|
Absolute calibration:
|
using pure ozone pressure
|
|
|
Uncertainty for absolute measurements at 50 mbar and 193 - 293K and path length 270 cm:
|
|
|
|
Systematic uncertainty
|
Ozone impurity: oxygen purity 99.995% leaks <0.1%. Accuracy of pressure sensors: 0.04% (0.02 mb) Offset of the internal temperature sensors: <0.3-0.5% (1K) Cell length: <0.07% (<1 mm) Total: <0.5%, depending on temperature
|
|
|
Statistical uncertainty
|
Ozone decay <1% Fluctuations of pressure: < 0.08% (<0.04mb) Fluctuations of temperature: <0.1 - 0.16% (<0.3K) Cooling efficiency: <0.3 - 1.6% (1-3K) Source intensity drift and white noise 0.1 - 70%, depending on spectral region
|
|
|
Concatenated spectra parameters:
|
|
|
|
Spectral regions:
|
Lightsource stability:
|
Optical density limits:
|
|
213–290 nm
|
0.5%
|
0.5-2
|
|
290–310 nm
|
2%
|
0.1-2
|
|
310–340 nm
|
1%
|
0.1-2
|
|
340–450 nm
|
1%
|
0.05-1
|
|
450–750 nm
|
0.2%
|
0.5-2
|
|
750–1100 nm
|
0.2%
|
0.001-0.1
|
|
|
|
|
|
column 1:
|
vacuum wavelength (nm)
|
|
|
column 2:
|
cross section (cm2/molecule)
|
293k
|
|
column 3:
|
cross section (cm2/molecule)
|
283k
|
|
column 4:
|
cross section (cm2/molecule)
|
273k
|
|
column 5:
|
cross section (cm2/molecule)
|
263k
|
|
column 6:
|
cross section (cm2/molecule)
|
253k
|
|
column 7:
|
cross section (cm2/molecule)
|
243k
|
|
column 8:
|
cross section (cm2/molecule)
|
233k
|
|
column 9:
|
cross section (cm2/molecule)
|
223k
|
|
column 10:
|
cross section (cm2/molecule)
|
213k
|
|
column 11:
|
cross section (cm2/molecule)
|
203k
|
|
column 12:
|
cross section (cm2/molecule)
|
193k
|